Magnetoencephalography (MEG)

MEG measures the extremely weak magnetic fields from neural currents (~ fT, a billionth of geomagnetic). Advantages: ms temporal resolution + better spatial localization than EEG (magnetic fields not distorted by skull). Needs SQUID + magnetic shielding (expensive). OPM-MEG (optically pumped magnetometer, 2018+) is a wearable revolution. Mainstay for source localization + oscillation research.

Difficulty: Advanced Prerequisites: EEG, Brain Rhythms


1. Principle

  • Neuronal postsynaptic currents → weak magnetic field (Biot-Savart)
  • Main source: pyramidal cells in sulcal walls (tangential currents, magnetically measurable)
  • Magnitude ~ 10-1000 fT (femtoTesla)
  • Detectors: SQUID (superconducting quantum interference device, needs liquid helium) or OPM

2. MEG vs EEG

MEG EEG
Measures Magnetic field Electric potential
Skull effect Almost none (magnetically transparent) Strong distortion/spread
Spatial localization Better (~ mm-cm) Poorer
Sensitive source Tangential (sulcal) Tangential+radial
Reference electrode Not needed (absolute) Needed (relative)
Portable Traditionally no (OPM yes) Yes

3. Inverse Problem (Source Localization)

  • Inferring intracranial current sources from surface fields = ill-posed inverse problem (no unique solution)
  • Methods:
  • Dipole fitting (equivalent current dipole)
  • Beamforming (LCMV)
  • Minimum norm estimate (MNE)
  • MUSIC
  • Needs head model (MRI co-registration) + prior constraints

4. PyTorch — Minimum Norm Inverse (simplified)

import torch

def minimum_norm_estimate(B, leadfield, reg=1e-3):
    """B: (n_sensors,) measured field; L: (n_sensors, n_sources) leadfield."""
    L = leadfield
    # MNE: J = L^T (L L^T + reg I)^-1 B  (Tikhonov-regularized)
    LLt = L @ L.t()
    inv = torch.linalg.solve(LLt + reg * torch.eye(LLt.size(0)), B)
    J = L.t() @ inv          # estimated source currents
    return J

5. OPM-MEG Revolution

  • Optically Pumped Magnetometer: no liquid helium needed, room temperature
  • Can be placed on scalp (closer to source → ↑ signal)
  • Wearable + allows head movement → natural behavior, children, clinical
  • Rapid development since 2018 (trend to replace SQUID)

6. Applications

  • Epilepsy: epileptogenic focus localization (presurgical) — main clinical use
  • Cognitive neuroscience: language, attention, memory spatiotemporal dynamics
  • Oscillations: gamma/beta (motor), theta — better localization than EEG
  • MEG-fMRI fusion: time (MEG) + space (fMRI)

7. Advantages Summary

  • ms temporal resolution (same as EEG)
  • Spatial localization better than EEG (fields not tissue-distorted)
  • No reference electrode, no electrode gel
  • Directly measures neural currents (unlike fMRI's indirect blood oxygen)

8. Limitations

  • Device very expensive + magnetic shielding room (traditional SQUID)
  • Insensitive to radial sources (gyral crown vs sulcal wall)
  • Weak signal for deep sources (rapid decay with distance)
  • Inverse problem non-unique (needs constraints + assumptions)
  • Motion-sensitive (traditional; OPM improves)

9. Relation to AI

  • Source localization = ill-posed inverse problem → deep learning reconstruction (like MRI reconstruction)
  • MEG decoding (decoding perception/language) + Transformer
  • Real-time BCI (OPM wearable → new possibilities)

10. Common Pitfalls

10.1 MEG = Magnetic EEG

Complementary but sensitive to different sources (MEG tangential, EEG also radial).

10.2 Spatial Localization Precise and Unique

Inverse problem non-unique; depends on head model + priors.

10.3 Measures All Neurons

Mainly tangential sulcal pyramidal cell synchronous activity; deep weak.

10.4 SQUID Is the Only Way

OPM-MEG room-temperature wearable is replacing it.

10.5 No MRI Needed

Source localization needs MRI anatomical co-registration (head model).



References

  1. Hämäläinen, M. et al. "Magnetoencephalography—theory, instrumentation, and applications." Rev Mod Phys, 1993.
  2. Baillet, S. "Magnetoencephalography for brain electrophysiology and imaging." Nat Neurosci, 2017.
  3. Boto, E. et al. "Moving magnetoencephalography towards real-world applications with a wearable system (OPM)." Nature, 2018.
  4. Gross, J. et al. "Good practice for conducting and reporting MEG research." NeuroImage, 2013.